Source Models, Internal Resistance & Loading
Real sources change terminal voltage or current when loaded. Thevenin and Norton source models make that interaction explicit and show why measuring open-circuit output is not enough to predict delivered power.
A source is useful only through the terminal behavior it presents to a load. A voltage reading with nothing connected identifies one point on that behavior, not the whole source. Internal resistance, current limiting, regulation, and nonlinear protection determine what happens once energy is drawn.
Linear source models #
The Thevenin form places an ideal voltage source in series with :
The Norton form places in parallel with the same resistance:
They are identical straight lines in the terminal - plane. The intercepts are:
| Condition | Measured quantity |
|---|---|
| Open circuit, | |
| Short circuit, | |
| Finite load |
For a safely testable linear source, two distinct operating points determine the line and hence . A short-circuit test is not required and may be destructive.
Loading error #
A voltmeter with input resistance is itself a load. Connected to a source with and , it reads
The familiar rule "use a high-impedance voltmeter" is therefore quantitative: the relative loading error is approximately when . An ammeter has the dual problem; its burden resistance changes the current being measured.
The source model must match the question. Use a constant-resistance equivalent for a narrow linear range; use a piecewise or dynamic model when the load crosses operating modes.
At AC, replace by the probe input impedance . The capacitive branch can dominate at high frequency even when the quoted DC input resistance is very large; this is the loading mechanism emphasized in the Tektronix probe guide.